EASE II
Whole medical range is available on
EASE II demand valve is a robust and compact device used by patients to self administer medical gas therapy. EASE II can be used to administer nitrous oxide and oxygen mixture (commonly known as N2O/O2) for pain relief or medical oxygen therapy. The EASE II demand valve is designed in a way that creates minimal breathing resistance to the patient and can deliver high flows when required.
EASE II O2 is recommended during diving accidents and cluster headaches therapy.
EASE II N2O/O2 is recommended during pain releave therapy.
FEATURES
- Deliver up to 300 litres per minute of gas
- Portable first stage regulator and cylinder version for immediate care and pre-hospital applications
- Conforms to global standards
- Low inspiratory effort demand valve
- Test/ Purge facility on the demand valve
- Easy grip handle and wrist strap
- Replaceable patient/bacterial filter
- Easy to clean and reassemble for cross infection protocol
- Hose fitted with probes by the national standards for connection into cylinder system or wall outlet
- Autoclavable at 134 °C
Art. Nr. | Gas Type | EASE II Variant Description |
0715306 | N2O/O2 | EASE II, 3 m hose, SS probe |
0715310 | O2 | EASE II, 3 m hose, SS probe |
ACCESSORIES
Art. Nr. | Description |
1032936 | Disposable mouthpiece |
1035575 | Breathing filter |
8848004 | Reusable universal size mask |
1024417 | Blue barrel carry bag |
5487294 | Mask adult size single use |
1032994 | Mask adult size single use - vanilla |
850500 | Expiration diverter single use for Ease II |
1570000SP | Scavenging hose |
TECHNICAL DATA
DEMAND VALVE
Gas: | O2, N2O/O2 |
Material: | Polycarbonate, silicone rubber, stainless steel |
Dimensiones: | 50 × 50 × 63 mm |
Weight: | 0,085 kg |
Gas supply: | Requirement 2,8 to 7,0 bar at >200 l/min |
Inspiratory resistance: | |
At 2,8 bar supply press.: |
Cracking 0,15-0,2 kPa
|
Expiratory resistance: | Cracking Zero At flow +0,35 kPa at 120 l/min |
Operating temperature: | -20°C to +60°C used Oxygen +5°C to +40°C used 50/50 N2O/O2 |
Storage temperature: | -30°C to +60°C |
Regulatory status: |
Complies with Medical Devices Directive 93/42/EEC Complies with BS 4272: Part 2: 1996
|
HOSE ASSEMBLY
Connection: | Probes by the national standards |
Working pressure: | 7 bar |
Burst pressure: | 44 bar |
Material: | PVC, anti-static in accordance with ISO 5359 |
Weight: | 0,5 kg (3 m length) |