ELITE
Group Number: 3422Whole medical range is available on
The innovative Elite oxygen conserving device conserves oxygen, and thus prolongs the service life of the cylinder.
- Totally pneumatic, requires no external power source
- On demand oxygen is delivered to meet patient needs, depending on exercise / effort
- Matching the delivery performance of constant flow oxygen over a range of flow rates
- Comprehensive range of international, high pressure connectors
- All brass high pressure components
- Easy to change between cylinders, medium pressure break ensures maximum user safety
- Easy to use, does not require complex set up, a simple, single connection of a cannula to the therapy outlet of the device is all that is required
Art. Nr. | Description |
2001635 | ELH-L-BULL-12F1 ELITEBOX |
2001636 | ELH-L-BENE-12F1 ELITEBOX |
2001637 | ELH-L-PI.O-12F1 ELITEBOX |
2001638 | ELH-L-DIN9-12F1 ELITEBOX |
2001639 | ELH-L-DN6O-12f1 ELITEBOX |
2001641 | ELV-L-E17-G1/4-12F1 |
2001642 | ELV-L-M18-G1/4-ELITE BOX |
2010294 | SABRE ELITE BS PROBE, BOXED |
TECHNICAL DATA
Input pressure: | 200 bar |
Pressure gauge: | for 200 bar filling pressure |
Pressure gauge reserve range: | 20 % of the filling presure |
Flow settings: | 1.0/ 1.2/ 1.5/ 2.0/ 2.5/ 3.0/ 3.5/ 4.0/ 4.5/ 5.0/ 5.5/ 6.0 l/min |
Pick-up pressure safety valve: | 7 bar |
Inlet Connector Type(s): | Full range high pressure |
Therapy outlet: | Universal firtree |
Service Intervals: | Elite - 5 Years |
Charging Adapter: | (Bullnose) Part Number: 2003558 |
Nasal Canula: | Part Numbers 2008202 (pack of 100) 2008203 (pack of 20) |